//(6) Override capability//
If the fluoroscopic x-ray field size is adjusted automatically as the [[SID]] or image receptor size is changed, a capability may be provided for overriding the automatic adjustment in case of system failure.

If it is so provided, a signal visible at the fluoroscopist's position shall indicate whenever the automatic field adjustment is overridden. Each such system failure override switch shall be clearly labeled as follows:
!!For X-ray Field Limitation System Failure