//(d)(2) Fluoroscopic equipment manufactured ''on or after May 19, 1995''//
* (i) Shall be equipped with AERC if operable at any combination of tube potential and current that results in an AKR greater than 44 mGy per minute (vice 5 R/min exposure rate) at the measurement point specified in [[1020.32(d)(3)|CFR 1020.32 Fluoroscopic equipment (d)(3) Measuring compliance]]. Provision for manual selection of technique factors may be provided.
* (ii) Shall not be operable at any combination of tube potential and current that will result in an AKR in excess of 88 mGy per minute (vice 10 R/min exposure rate) at the measurement point specified in [[1020.32(d)(3)|CFR 1020.32 Fluoroscopic equipment (d)(3) Measuring compliance]], except as specified below (clause 1020.32(d)(2)(iii))
** (A) For equipment manufactured prior to June 10, 2006, during the recording of images from a fluoroscopic image receptor using photographic film or a video camera when the x-ray source is operated in a pulsed mode.
** (B) For equipment manufactured on or after June 10, 2006, during the recording of images from the fluoroscopic image receptor for the purpose of providing the user with a recorded image(s) after termination of the exposure. Such recording does not include images resulting from a last-image-hold feature that are not recorded.
** (C) When a mode of operation has an [[optional high-level control and the control is activated|CFR 1020.32 Fluoroscopic equipment (d)(2) Optional clause for high-level control]]